39
10601034 Rev. B
LENS Integrated System
Operations/Service Manual
Troubleshooting
Problem
Try this
Probable cause
• Contact the local system administrator.
• Ensure that the iPad is within visible range (within 150 ft) of
the control unit.
If this works but persists within the O.R., contact technical
support.
The iPad is out of range of the control unit
wireless network
• Check Wi-Fi antenna is connected on back of system
Damaged wifi antenna
Cannot capture images/video to
the App
• Ensure connection to the control unit Wi-Fi network.
iPad is not connected to the control unit
• Check that Wi-Fi antenna is connected on to the back of
control unit.
If yes, replace the Wi-Fi antenna.
If no, connect a Wi-Fi antenna.
The Wi-Fi antenna was damaged or missing
Cannot capture images/video to
the USB storage device
• Refer to the compatible devices in this manual and attempt
to reconnect.
Call tech support
• Inspect for damage to the USB ports on the system (use the
alternate port).
Physical damage to port
Camera color is not correct
• Perform a white balance per instructions in IFU.
White balance was not successfully
completed
• Restore factory defaults from system configuration menu
(note this will erase all current settings and restore to
factory).
Color phase or hue were incorrectly
customized in the procedures
• Unplug and plug in the camera head and perform white
balance.
If this works and is repeatable, call tech
support
System AC Mains LED is not lit
• Check AC mains switch on back of unit and ensure that the
proper mains cable and outlet are used.
Mains switch should be enabled on (
I )
• Replace the fuse.
Fuse was blown
If this problem persists contact local facilities, then call technical
support.
No video displayed on the
monitor or 660HD System
• Check cables (correct cable, actual connection, integrity of
the cables (cracks, breaks, etc.).
Bad cables. Use the correct functional cables.
• Bring up the color bar from MAIN MENU and check all video
outputs on the monitor.
Bad camera head. Try another camera head.
Check the video output format in Advanced Settings menu of
the control unit and ensure that settings are compatible with
monitor or 660HD Image Management System. Refer to the
Advanced Settings section of this Instructions for Use for more
information.
Note: The 660HD System accepts only 1080i HD-SDI output.)
Mismatched video format setting
Arthroscope is too hot
Ensure that the proper light guide diameter is in use for the
current operative procedure.
Incorrect light guide
Ensure that the light guide to light post connection is secure.
Faulty connection
Turn down the brighness of the control unit and contact Smith &
Nephew Customer Service.
Not known