Instruction Manual OilGuard Ex M
10241E/2
39
Action
Display (example)
Remarks
2.
3 x
p
> Adju. <
Retain
3.
2
Adju.
> Retain <
4.
p
Adju.
> Adapt <
5.
2
Adju.
running ...
6.
Wait until the
photometer has
determined the
calibration factors
and the display
changes.
- Measure –
X.XXX FLU
Display of "raw“ check
values without reading
adjustments.
If “Tolerance exceed”
appears in the display,
consult the following
section.
7.
After successful recalibration of the instrument, rebuild to the initial state.
If the nominal and actual values differ from each other too greatly, check the
following points:
Is the serial number of the photometer the same as the one on the control
unit?
Does the nominal value match the one on the control unit?
Is the control unit clean and undamaged?
Are the measuring cell windows clean?
Has the minimum warm-up time of two hours been observed?
If this does not help, please contact customer service (
→
Section 6.4).
What to do if value
is “out of tolerance”