Instruction Manual OilGuard Ex M
20
10241E/2
3.4
Initial startup
For the initial startup, proceed according to the following table. In the event of
faults, please refer to Section 6.
Action Remarks
1.
Make sure that the photometer is
correctly mounted and connected.
→
Section 3.1 and 3.2
2.
Close the front door.
Place the seal on the front door.
Make a note of the seal number in
the maintenance protocol.
→
Section 0
→
Section 11
3.
Open the purge air line and set the
pressure to 2 bar using the
regulator screw (A).
4.
Check the sampling system for
correct sample supply and open the
sample feed.
The following criteria must be
observed:
1.
Is the sample flow rate stabile?
2.
Is the sample flow degassed?
5.
Connect the power supply to the
instrument.
The instrument is first purged about
2 min., and then it switches on and
a reading appears on the display.
6.
Check the Data logger to ensure it
is functioning properly.
→
Section 5.6
7.
Set the language for your region
(
→
Section 4.5).
The menu text now appears in the
selected language.
8.
Protect your settings against
unauthorized access with an
access code (
→
Section 4.7).
If you do not require an access code,
you can skip this step.
First commissioning