Establish the electrical connections in the following sequence:
Terminals Meaning
Remarks
4 - 5 - 6
Relay output 1
7 - 8 - 9
Relay output 2
→
Section 4.6
→
Reference Handbook
11, 12, 20,
21
data logger
→
Section 3.2.6
1.
18 - 19
Output signal
0/4 .. 20 mA, max. load 600
Ω
If unused, these terminals must
be short-circuited with a jumper.
10, 14
State signal from
bilge separator
The state of the bilge separator is
polled at input MRIn 2
0
.
→
Reference Handbook
10, 16a
Flow rate monitor
If unused, these terminals must
be short-circuited with a jumper.
20 - 21
Power supply to bus
coupler
→
Reference Handbook
2.
24 - 25
Data connection to
the bus coupler
→
Reference Handbook
3.2.6
Connecting the OilGuard data logger
The data logger is not explosion protected. It may therefore be used only in
the “non-Ex zone”.
Pos. Name
1 Connection
terminals
2 Backup
battery
Figure 9: Position of the connection terminals in the OilGuard data logger
To connect the data logger, use a connection cable with a minimum wire
cross-section of 4 x 0.25 mm
2
.
Connecting the
photometer
Position of the
connection terminals
in the data logger