Replace inlet tube
Action
1.
Remove the sample feed and return line
and let the measuring cell run empty.
2.
Remove the lock nut by turning it
counter-clockwise (bayonet connector)
and then pull out the inlet tube.
3.
Insert the new inlet tube (A) and mount
the lock nut (B).
Fix the inlet tube in place by turning the
lock nut clockwise (B).
4.
Re-attach the sample feed and return line.
Replace inlet tube