Instruction Manual OilGuard Ex M
24
10241E/2
4.5
Setting the language
You can set the language of the menus and messages to your region as
follows:
Action
Display (example)
Remarks
1.
Activate service
mode
* RECALIBR. *
* *
Section 4.4
2.
3 x
n
* CONFIGURATION*
3.
p
> Language <
German
4.
2
Language
> German <
Activate editing mode.
5.
Select language:
q
/
p
Language
> ... <
6.
2
> Language <
...
Confirm selection.
7.
q
+
p
(simultaneously)
24.0 ppm 1
Unit in normal operation.
4.6
Setting the relay functions
The photometer has two relay outputs; only the settings of relay 2 can be
changed by the operator (
→
Section 3.2.5
).
The factory settings of the relays are as follows (CAPITAL letter functions are
activated, e.g. al
→
AL):
Abbrev. Meaning Setting Remarks
LI Limit
exceeded
fixed
After the photometer is
switched on, the display
shows a limit exceed during
the 10-minute warm-up
phase.
AL
Alarm (fault has occurred)
SE
Instrument in service mode
AD
Adjustment in progress
IN Relay
inverted
fixed
Language
Configure relay
functions
Relay 1
Not configurable