Instruction Manual OilGuard Ex M
10241E/2
25
Abbrev. Meaning
Setting
li Limit
exceeded
al
Alarm (fault has occurred)
se
Instrument in service mode
ad
Adjustment in progress
Changeable
IN Relay
inverted
fixed
Action Display
(example)
Remarks
1.
Activate service
mode
* RECALIBRATION*
* *
Section 4.4
2.
4 x
n
* RELAY *
* *
3.
4 x
p
> Relay 2 <
li al se ad IN
Configure relay 2
4.
2
Relay 2
>li al se ad IN<
Activate editing mode
li= Limit
2
exceeded
al =
Alarm (fault
occured)
se = Instrument in service
mode
ad = Adjustment in
progress
5.
Assign functions:
o
/
n
Function
On/Off
q
/
p
Change
function
Relay 2
>LI al se ad IN<
in =
Relay inverted
6.
2
> Relay 2 <
LI al se ad IN
Confirm selection
7.
p
> R2 on delay <
00000 s
Relay 2 switch-on delay
→
Reference Handbook
8.
Change value:
o
/
n
Change
numerical
value
q
/
p
Change
digit
pos.
R2 on delay
> 00050 s <
9.
2
Confirm value
> R2 on delay <
00050 s
10.
p
> R2 off delay <
00000 s
Relay 2 switch-off delay
→
Reference Handbook
Relay 2
Configurable
Programming relay 2