Instruction Manual OilGuard Ex M
10241E/2
19
3.3
Check connection between PC and OilGuard
data logger
The connection between the OilGuard data logger and the PC can be checked
in the following way:
Action Remarks
1.
Start up the HOBOware software and read out
the "Logger state" by clicking button (A).
The “State” page then appears (
→
Figure 13 )
2.
Check the entries in accordance with the
following specifications:
-
At "Current Status" (2), "Launched, Logging"
must be displayed.
-
"Memory used" (1) must not be full.
-
The displays of "Current Readings"(3), "Limit"
and "Error status" must correspond to the
current states.
→
Figure 13
Pos. Name
1 Memory
used
(free memory
capacity)
2 Current
Status
3 Current
Readings
Figure 13 State of the logger
Check connection
between Data logger
and PC
Information about
the status of the
logger