SDS5000XSeries Digital Oscilloscope Service Manual
74/ 90
WWW.SIGLENT.COM
8.5.1
Acquisition Board SDY8.007.163
8.5.1.1
Acquisition Board Drawing
It is recommended to remove the acquisition board from the metal
frame and re-assemble the power supply module to the metal frame
for easy access to the test points and safety considerations. Follow
the instructions in chapter Disassembly Procedures.
Reconnect the power cable to the acquisition board after you have
confirmed that the power supply module is in good condition.
Power up the acquisition board by pushing the front panel power
button.
In most cases, you can hear the relays click if the acquisition board
power up successfully. If you cannot hear the click or there is no sign
that the acquisition board is powering up, go toTo Check the Power
Supply Modulesection to check the stand-by power and the main
power.
If the power or the signal under test is abnormal, replace the