SDS5000XSeries Digital Oscilloscope Service Manual
79/ 90
35
VCC-13V5
TP48
-13.5
V
𝐷𝐶
±
400
mV
𝐷𝐶
36
VCC-10V5_FE
TP143
-10.5
V
𝐷𝐶
±
300
mV
𝐷𝐶
37
VCC-12V_PROBUS
TP130
-12
V
𝐷𝐶
±
360
mV
𝐷𝐶
38
VCC12V_PROBUS
TP49
+12
V
𝐷𝐶
±
3600
mV
𝐷𝐶
39
VCC1V8_FE
TP134
+1.8
V
𝐷𝐶
±
50
mV
𝐷𝐶
40
VCC2V5_FE
TP133
+2.5
V
𝐷𝐶
±
100
mV
𝐷𝐶
8.5.1.3
Check the clock
Check the oscillator X2.Use an oscilloscope to measure the clock signal
using the method below.
Figure 27
Test points for clock
Table 16
Clock parameters of the Acquisition System
Test Point
Clock Domain
Level
Description
TP44
PLL Reference Clock LVCMOS33 Typical 10 MHz