SDS5000XSeries Digital Oscilloscope Service Manual
88/ 90
WWW.SIGLENT.COM
Connect the LCD cable to the acquisition board.
Power up the acquisition board by pushing the front panel power
button.
8.7.1
To Check the LCD Power Supply
Find and measure those test points using a multi-meter.
Figure 35
Test points for the LCD power supplies
Table 21
Voltage parameters of the LCD power supplies
Net Name Test Point
Tolerance
12V6_LCD
L71
+12.6
V
𝐷𝐶
±
10%
-6V5_LCD
L72
-6.5
V
𝐷𝐶
±
10%
VCC3V3
J32 PIN
2/4/6
3.3
V
𝐷𝐶
±
10%
If the power supply is abnormal, replace the acquisition board.
8.7.2
To Check the LCD Signal
Locate connector J32, use an oscilloscope to measure the signal using
the method below.