SDS5000XSeries Digital Oscilloscope Service Manual
86/ 90
WWW.SIGLENT.COM
Figure 33
Test points for the processor power supplies
Table 19
Voltage parameters of the processor power supplies
Net Name
Test Point Tolerance
DVCC1V0_ZYNQ TP1
+1.0
V
𝐷𝐶
±
50
mV
𝐷𝐶
DVCC1V8_ZYNQ TP2
+1.8
V
𝐷𝐶
±
50
mV
𝐷𝐶
DVCC1V5_ZYNQ TP3
+1.5
V
𝐷𝐶
±
50
mV
𝐷𝐶
DVCC3V3_ZYNQ TP4
+3.3
V
𝐷𝐶
±
100
mV
𝐷𝐶
DVCC2V5_ZYNQ C129
+2.5
V
𝐷𝐶
±
100
mV
𝐷𝐶
8.6.2
Check the clock
Check the three oscillators X1, X2, and X3.Use an oscilloscope to
measure the clock signal using the method below.