6.3.3
To Remove the Power Supply Module and Fan
Figure 16
To remove the power supply module and Fan
Steps:
1.
Remove the tenPM3*6 screws outside the metal rear cover.
2.
Remove the metal rear cover.
3.
Unplug the connecting cable from the power module.
4.
Remove the ten screws on the power supply module (Six PM3*6
screws, four PMW3*6 screws).
5.
Remove the power supply module.
6.
Remove the four KM3*32 screws and nuts on the fan and remove the
fan.