SDS5000XSeries Digital Oscilloscope Service Manual
32/ 90
WWW.SIGLENT.COM
EXT
PC
CH
1
CH
2
SDS5000X
GPIB
LAN
9500B
CH1
BNC CABLE
Figure 12
Connecting test instruments for Ext Trigger Sensitivity
PC
CH1 CH2
SDS5000X
GPIB
LAN
9500B
CH1
CH2
CH3 CH4
BNC Cable
Figure 13
Connecting test instruments for Channel Skew
Verify Test Results
To verify whether or not a test passes (i.e. whether the readings are
within the appropriate limits), it is necessary to record the readings in the
Performance Test located in the Test Record.
Self-Calibration
The Self Calibration procedure is described in sectionSelf-Calibration. If
the environmental temperature changes by more than 5
℃
, the Self