SDS5000XSeries Digital Oscilloscope Service Manual
54/ 90
WWW.SIGLENT.COM
Channel Skew Limit: abs(T@
L) ≤ 1
50 ps
Steps:
1.
Connect two selected channels of the oscilloscope to the 9500B
shown as in
2.
Press the ‘Aux’ key at the right of the ‘OSCILLOSCOPE CALIBRATOR’
panel of the 9500B and select Skew function.
3.
Set impedance of the 9500B to 50 Ω, frequency to 1
kHz, and
amplitude to 0.6 V.
4.
Set the oscilloscope Volts/Divto 100mV/div.
5.
Set the oscilloscope Sec/Divto 2ns/div, trigger source is the selected
channel.
6.
Press theMeasurebutton on the front panel of the oscilloscope to
display Skew measurement.
7.
Check if the result is in the limited range above.
8.
Disconnect the test connection.
9.
Check other channels in the same way as steps 1 to 7.
5.13
To Verify Input Impedance
This test checks the input impedance of all analog channels and the ext
trigger channel with different coupling modes and vertical scales (100
mV/div, 200 mV/div and 2 V/div).