SDS5000XSeries Digital Oscilloscope Service Manual
49/ 90
≤
234M
(-3,1) dB
≤
351M
(-5,1) dB
≤
501M
(-44,3) dB
5.10
To Verify Bandwidth Limit
This test checks the bandwidth limit of all the analog channels. The test
is similar to the bandwidth test except that the bandwidth limit is turned
on. In the test, the impedance of 9500B should be set to 50
Ω.Test the
bandwidth at 100mV/div, 200 mV/div
Steps:
1.
Connect channel 1 of the oscilloscope to the 9500B via the active
head, as shown in
2.
Set the oscilloscope Volts/Div to 100mV/div, the Sec/Div to 50ms/div,
and set the BW Limit button to 20MHz.
3.
Press the Measure button on the front panel of the oscilloscope to
display Peak-Peak measurement.
4.
Set the amplitude of the 9500B to 600m
V, and impedance to 50Ω
,
and set the impedance of the oscilloscope channels to
50 Ω
.
5.
Set frequency of the 9500B to the frequency in
Table 9
Frequency points in the test