SDS5000XSeries Digital Oscilloscope Service Manual
63/ 90
6.3.5
To Remove the Keyboard and Probe Interface Board
Figure 18
To removethe keyboard and probe interface board.
Steps:
1.
Remove the seven PM3*6 screws on the keyboard.
2.
Disconnect the cable between the probe interface board and the
keyboard.
3.
Remove the keyboard.
4.
Disconnect the cable between the mainboard and probe interface
board.
5.
Remove the two PM3*6 screws on the probe interface board.
6.
Remove the probe interface board.