SDS5000XSeries Digital Oscilloscope Service Manual
55/ 90
Table 12
Input Impedance data
Channel
Coupling and
impedance type
Impedance Range
CH
AC/1 MΩ
0.1V/div
1176000~1224000 Ω
0.2V/div,2
V/div
980000~1020000 Ω
DC/1 MΩ
980000~1020000 Ω
DC(AC)/50
Ω
49.15~50.85
Ext
1 MΩ
980000~1020000 Ω
50
Ω
49.15~50.85
Steps:
1.
Connect the selected channel of the oscilloscope to the 9500B, as
shown in
2.
Set the oscilloscope Volts/Div to 100 mV/div, coupling mode of the
channel to AC.
3.
Press the ‘Aux’ key at the right of the ‘OSCILLOSCOPE CALIBRATOR’
panel of the 9500B and select the Load Resistance Measurement
Function. Record the reading displays on the screen of the 9500B.
4.
Set coupling mode of the channel to DC. Record the reading displays
on the screen of the 9500B.
5.
Check if the impedance is within the specified range in
6.
Test at 2 V/div and 200mV/div, using the same procedure as steps 2