SDS5000XSeries Digital Oscilloscope Service Manual
26/ 90
WWW.SIGLENT.COM
4.2.4
Pass/Fail out Test
To test the Mask Test function and signal output by viewing on
anotheroscilloscope.
Tools:
A second oscilloscope
Two BNC cables
Steps:
1.
Turn on the SDS5000X oscilloscope.
2.
Enable Channel 1.
3.
Perform
Analysis>MaskTest to open the Mask Test dialog box, and
touch
Mask Setup>Create Mask to set the mask. Set the items as
below:
Table 1 Mask Test Settings
Parameter
Setting
Enable to Test
On
Source
CH1
Type
All in
Display Info
On
Mask Setting
Mask X
Mask Y
0.32
0.32