SDS5000XSeries Digital Oscilloscope Service Manual
83/ 90
Figure 31
Test points for the Acquisition board power supplies (continued)
Table 17
Voltage parameters of theAcquisition System power supplies
Number
Net Name
Test Point Tolerance
1
VCC12V6
J24
+12.6
V
𝐷𝐶
±
400
mV
𝐷𝐶
2
VCC1V0A
C1046
+1.0
V
𝐷𝐶
±
50
mV
𝐷𝐶
3
VCC1V0B
C1058
+1.0
V
𝐷𝐶
±
50
mV
𝐷𝐶
4
VCC1V8
TP73
+1.8
V
𝐷𝐶
±
50
mV
𝐷𝐶
5
VCC1V5A
TP75
+1.5
V
𝐷𝐶
±
50
mV
𝐷𝐶
6
VCC1V5B
TP77
+1.5
V
𝐷𝐶
±
50
mV
𝐷𝐶
7
VCC2V5
TP83
+2.5
V
𝐷𝐶
±
100
mV
𝐷𝐶
8
VCC3V3
TP85
+3.3
V
𝐷𝐶
±
100
mV
𝐷𝐶
9
VCC5V
TP81
+5.0
V
𝐷𝐶
±
150
mV
𝐷𝐶
10
VCC1V2
TP79
+1.3
V
𝐷𝐶
±
50
mV
𝐷𝐶
11
VREF_DDR_A
TP63
+0.75
V
𝐷𝐶
±
22.5
mV
𝐷𝐶
12
VREF_DDR_B
TP65
+0.75
V
𝐷𝐶
±
22.5
mV
𝐷𝐶