SDS5000XSeries Digital Oscilloscope Service Manual
53/ 90
6.
Check if the frequency is within the specified trigger range listed in
Table 10.
7.
Set trigger slope to Negative, repeat steps 4 to 6.
8.
Set frequency of the 9500Bto the oscilloscope’s bandwidt
h frequency,
amplitude to 300 mV. Repeat steps 3 to 7. In step 5, the range
changes to [-150 mV, 150 mV].
9.
Set frequency of the 9500Bto 10 MHz, amplitude to 1 V.
10.
Set the trigger source of the oscilloscope to EXT/5, trigger slope to
Positive.
11.
Press Trigger Level knob to set the level to the center of the
waveform.
12.
Adjust the trigger level within the range of [-500mV, 500mV] to
achieve a stable waveform and then record the
hardware-measuredvalue which is on the top right of the screen.
13.
Check if the frequency is within the specified range listed in Table 13.
14.
Set the trigger slope to Negative, repeat steps 11 to 13.
15.
Set the frequency of the 9500Bto the oscilloscope’s bandwidth
frequency, amplitude to 1.5 V. Repeat steps 11 to 14.In step 6, the
range changes to [-750 mV, - 750 mV].
16.
Disconnect the test connection.
5.12
To Verify Channel Skew
This test checks the skewbetween two analog channels. In this test, the
impedance of both the 9500B and the oscilloscope should be set to 50
Ω.