Analog Electronic Modules
12-25
ET 200S Distributed I/O System
EWA-4NEB 780602402-12
•
Relevant parameters for interface module
Parameters
Value
Reference junction slot
3
Reference junction input
RTD on channel 0
•
Relevant parameters for 2AI RTD and 2AI TC:
Slot
Parameters
Value
8 (2AI RTD)
Measurement type / range I0
RTD-4L Pt100 terminal
5 (2AI TC)
Reference junction I0
RTD
Reference junction E1
No
Reference junction number
1
Measurement type / range I0
TC-EL type...
Measurement type / range I1
(any)
8 (2AI TC)
Reference junction I0
RTD
Reference junction E1
RTD
Reference junction number
1
Measurement type / range I0
TC-EL type...
Measurement type / range I1
TC-EL type...
11 (2AI TC)
Reference junction I0
No
Reference junction E1
RTD
Reference junction number
1
Measurement type / range I0
(any)
Measurement type / range I1
TC-EL type...
Non-insulated thermocouples
If you use non-insulated thermocouples, make sure that you adhere to the
permitted common mode voltage.
Содержание SIMATIC ET 200S
Страница 42: ...Product Overview 1 16 ET 200S Distributed I O System EWA 4NEB 780602402 12 ...
Страница 64: ...Brief Instructions on Commissioning ET 200S 2 22 ET 200S Distributed I O System EWA 4NEB 780602402 12 ...
Страница 112: ...Configuration Options 3 48 ET 200S Distributed I O System EWA 4NEB 780602402 12 ...
Страница 232: ...Commissioning and Diagnostics 6 72 ET 200S Distributed I O System EWA 4NEB 780602402 12 ...
Страница 244: ...General Technical Specifications 7 12 ET 200S Distributed I O System EWA 4NEB 780602402 12 ...
Страница 274: ...Interface Modules 8 30 ET 200S Distributed I O System EWA 4NEB 780602402 12 ...
Страница 538: ...4 IQ SENSE 13 18 ET 200S Distributed I O System EWA 4NEB 780602402 12 ...
Страница 564: ...Address Space of the Input and Outputs of the ET 200S C 4 ET 200S Distributed I O System EWA 4NEB 780602402 12 ...
Страница 574: ...Response Times D 10 ET 200S Distributed I O System EWA 4NEB 780602402 12 ...
Страница 580: ...Special Measures for Interference Free Operation F 4 ET 200S Distributed I O System EWA 4NEB 780602402 12 ...
Страница 600: ...Product Information on the Use of subassemblies modules in a Zone 2 Hazardous Area 2 A5E00352937 02 ...