Prerequisites / General Information
1 - 3
Siemens AG
RXL2-120.815.01
Page 3 of 6
LITHOSTAR Multiline
Medical Solutions
Rev. 02
02.05
CS PS 24
System manual
Protective Measures
1
•
Adhere to the protective measures contained of the LITHOSTAR Multiline System
binder.
•
Set the main system switch to
OFF
to switch off power to all components (generator and
connected equipment).
•
Switch
OFF
(S1) SS on the D 100 board (Fig. 1, no inverter power) to avoid inadvertent
release of high voltage or radiation.
•
Install or remove assemblies only with the generator switched off and according to ESD
guidelines.
Preliminary Steps
1
•
Set switch (S1) on board D 100 to SS
OFF
(Fig. 1).
Tests and adjustments performed with radiation ON are identified
by the radiation warning symbol
.
Radiation protection must be worn during these types of adjust-
ments (refer to ARTD part 2).
Voltage is still present in transformer T1 and power-on circuit
D160 even after the generator is switched off (see wiring dia-
grams X2206-10, X2169-10).
Following generator switch-off a voltage of approximately 600 V
DC is still present for the converter! LEDs V35 and V36 on D110
(Fig. 1) indicate this (X2206-16, X2169-10) and LED V89 (X2206-19,
X2168-019) lights up.
The voltage drops to 0 V in approx. 1
1
/
2
minutes; the LED’s go out
when the voltage reaches approximately 30 V.
NOTICE
WARNING