LITHOSTAR Multiline
RXL2-120.815.01
Page 4 of 6
Siemens AG
System manual
Rev. 02
02.05
CS PS 24
Medical Solutions
3 - 4
Shock Wave System
•
Fill the coupling bellows with water by activating the S1 key on board W12 (Fig. 1)
several times.
•
Remove any air that has collected in the coupling bellows using a syringe without a
needle.
•
Set switch S2 to the EMPTY COUPL. BELLOWS position (Fig. 1) (valve Y3 is open).
•
Connect the funnel to fill the water system (2/Fig. 2).
•
Fill the water tank so that the water level is between the marks on the upper sight glass
(1/Fig. 2).
•
Remove the funnel (release the quick-connector on the water tank) (2/Fig. 2).
•
Set the S2 switch on board W12 to the STANDARD MODE position (Fig. 1) (Y3 valve
open).
•
Fill the coupling bellows with water using the S1 button on W12 board (Fig. 1).
•
Remove air bubbles in the coupling bellows using a syringe (without a needle) (Fig. 3).
•
Set the S2 switch on board W12 to the EMPTY COUPLING BELLOWS position (Fig. 1)
(Y3 valve is open).
Fig. 3
The level in the water tank must not drop below the mid range of
the lower sight glass (3/Fig.2).
Fill the water tank only when the coupling bellows has completely
collapsed (a minimal amount of water remains in the coupling bel-
lows).
The level in the water tank must not drop below the mid range of
the lower sight glass (3/Fig. 2).
NOTICE
NOTICE
NOTICE