LITHOSTAR Multiline
RXL2-120.815.01
Page 2 of 4
Siemens AG
System manual
Rev. 02
02.05
CS PS 24
Medical Solutions
2 - 2
General Start-Up
•
Switch the system breaker
OFF.
•
Remove the internal line impedance meter and test the other phases.
•
Record the measured value in the table below or in the Multiline test report.
•
To reach full generator power, the measured internal power line impedance may not
exceed the following value.
•
Switch the system breaker
OFF
.
•
Reinstall the knife fuses.
•
Reinstall the cover over the knife fuses.
U
Line
max. R
Line
for POLYDOROS LX 50
Measured
400 V
±
10%
0.15 Ohm
max. R
Line
for POLYDOROS SX 65
400 V
±
10%
0.14 Ohm
max. R
Line
POLYDOROS SX 80
400 V
±
10%
0.11 Ohm