LITHOSTAR Multiline
RXL2-120.815.01
Page 2 of 2
Siemens AG
System manual
Rev. 02
02.05
CS PS 24
Medical Solutions
7 - 2
Start-Up and Tests of the Options
Multispot 2000
7
•
Select the test image
0C
(refer to Setting of Software Parameters RXL2-120.032...).
•
Switch on the MULTISPOT 2000.
•
Insert cassette with film. Refer to the operating instructions for MULTISPOT operation.
•
Release a hardcopy.
•
Check the Multispot functions.
•
Check the adjustment of the density according to the instructions RX57-091.032... .
•
Reset the test image to
00
.
Setting the Time on the ECG Monitor
7
•
Set the time in accordance with the Patient Monitor Instructions.
Start-up of Sonoline Adara
7
•
Turn the unit on.
•
Refer to Chapter 4 System set-up "General system settings" in the SONOLINE Adara
Operating Instructions for setting the date, time, language and hospital name.
•
Attach the universal adjustment device
•
Adjust the centering according to the "Repair instructions ultrasound localization
RXL2-120.091.10...".
•
Remove the universal adjustment device (the universal adjustment device must not be
subjected to shock waves).
Prerequisite:
Fluoroscopic image (LIH) is displayed on the right screen of the dual
monitor.
•
Shockwave release test
"
test with ECG triggering
"
test with respiratory triggering
"
test with respiratory and ECG triggering