Start-Up and Tests of the Options
7
ECG Triggering
7
•
Select ECG triggering via the Operating PC.
Refer to the LITHOSTAR Multiline operating instructions.
•
Release shock waves and check their synchronization to the pulse rate.
•
Deselect ECG triggering.
Respiratory Triggering
7
•
Connect the respiration belt to the unit.
•
Select respiratory triggering at the Operating PC.
•
Simulate respiration.
•
Wait for the learning phase to take place, i.e., several respiration cycles.
•
Release shock waves and check their synchronization to the respiration rate.
•
Repeat the test with the second connection.
Respiratory and ECG triggering
7
•
Set up triggering for both as described above and select both at the same time.
Release shock waves and check their synchronization to the pulse and respiration rates
(refer above).
7
Diamentor
•
The Diamentor is calibrated at the factory.
•
Checking procedures and recalibration are described in the instructions accompanying
the Diamentor.
DICOM Connect
7
•
Perform the start-up for DICOM Connect according to the instructions RXL2-120.814.02.
Prerequisite:
The shock wave head is in the therapy position and the coupling pres-
sure is on. ECG electrodes are attached to the patient.
Normally, each ECG pulse triggers a shock wave. At higher pulse
rates, only every other pulse triggers a shock wave.
Several shock waves are triggered in each expiration phase.
DICOM can be installed in the LITHOSTAR Multiline only if the
MULTISPOT is not installed.