Start-Up of Multiline System
4 - 5
Siemens AG
RXL2-120.815.01
Page 5 of 8
LITHOSTAR Multiline
Medical Solutions
Rev. 02
02.05
CS PS 24
System manual
•
Confirm the message "Backup was successful" with OK.
•
Exit the service program with System quit.
•
Enter the desired text for "Working on:" e.g."Start-up" and then click "Put to unit".
•
Switch the system off and then on again.
Checking kV Accuracy
4
(Only if the cassette option is present)
•
Install the cassette holder on the Multiline and insert an empty cassette (35 x 43).
•
Select 77 kV, 32 mAs and 100 % power for both
and
via the Operating PC.
•
Release an exposure for each and check the kV value on the oscilloscope.
•
Enter the measurement value in the generator test protocol.
After each exposure, remove the cassette and reinsert it.
Use a kV meter or the NOMEX for acceptance tests.
The corresponding instructions and test conditions can be found in the Operating
Instructions.
Exposure Indicators on the Operating PC
4
(only if the cassette option is present)
•
Select 77 kV, 100 mAs,
80 % power at the service PC.
•
Release an exposure:
- During the exposure (radiation ON), the radiation symbol must appear on the service
PC.
- The audible signal of the Operating PC must be activated for the duration of the
exposure (beeper signal).
•
Remove the cassette after each exposure and reinsert it.
Checking the mAs Values
4
(only if the cassette option is present)
•
Select the following values at the Operating PC:
, 77 kV, 80 mAs, 100 % power.
•
Release an exposure and check the mAs value on the mAs meter. After each exposure,
remove the cassette and reinsert it.
•
Enter the measurement value in the generator test protocol.
Checking the Maximum Generator Power
4
(only if the cassette option is present)
•
Select
, 102 kV, 100 % power at the Operating PC and select 28 mAs.
•
Select 2V/div for kV
act
and 1V/div for mA
act
on the oscilloscope.
•
Release an exposure and measure the kV and mA with the oscilloscope.
p
p