Checking the Dose Rate
6 - 15
Siemens AG
RXL2-120.815.01
Page 15 of 18
LITHOSTAR Multiline
Medical Solutions
Rev. 02
02.05
CS PS 24
System manual
Message line:
Start fluoroscopy with
ON
Message line:
Turn D100 R1 in the camera adapter to the right.
•
Set the adjustment in the camera adapter to +1 (+1/4 EP) with R1 on board D100.
•
If this range is not adequate, change over to switch S2 on board D100 in the camera
adapter and repeat the adjustment.
•
Remove the measurement chamber.
•
Leave 1.2 mm Cu filtration and exit the window with OK; the window PDA Preadjust
appears.
•
With fluoro
ON
, check whether the value 0 +1 is attained in the field "Difference of
expected to actual value".
•
Otherwise, adjust potentiometer R1 on D100 until the adjustment is successful.
7
S2
R1
negative value
−−>
CCW