LITHOSTAR Multiline
RXL2-120.815.01
Page 4 of 8
Siemens AG
System manual
Rev. 02
02.05
CS PS 24
Medical Solutions
4 - 4
Start-Up of Multiline System
•
Confirm the message "Please close the collimator" with OK.
•
Select "Fluoroscopy" in the Adjustment menu.
•
Switch on fluoroscopy (i.e. click "on" using the mouse). After approximately 10 minutes,
fluoroscopy switches off automatically.
•
Click OK.
•
Switch on "Fluoroscopy" in the Pre-Heating field (click ON with the mouse).
Fluoroscopy switches off automatically.
•
Click OK after the fluoro has switched off.
•
Switch on "Correction Exposure" in the Learn Filament field (i.e. click ON using the
mouse). The correction is determined for Load 10 %.
•
Release another exposure after the correction. The correction for Load 60 %
(large focus) and 80 % (small focus) is performed.
•
Click OK.
•
Release an exposure in the Learn Filament Pushcurrent field (i.e. click ON with the
mouse).
•
Continue to release exposures until the message: "Release next exposure" disappears.
•
Click OK.
•
The query appears: Next focus (click cancel if second focus is set already). If you need to
perform the adjustment for the second focus, click OK in the ADJ menu.
Select warm-up and repeat the procedure.
•
Click OK after fluoro switches off.
•
Switch on an exposure in the Learn Filament Correction field (click ON using the mouse).
The correction is determined for Load 10%.
•
Release another exposure after the correction. The correction is determined for Load
60 % (large focus).
•
Click OK.
•
Release an exposure in the Learn Filament Pushcurrent field (i.e. click ON using the
mouse).
•
Release additional exposures until the message: "Please wait...exposure" no longer
appears.
•
Click OK.
•
Click CTRL + X in the System Quit menu.
•
Reset switch S3/X44.
•
Confirm the subsequent message "Set the S3/X44 switch ..." with OK. The green LED
V90 on board D100 goes out (refer to Fig. 2).
•
Click OK.
•
Click "Backup to Disk" in the Data menu.
•
Insert the Site data floppy.
•
Click OK.