Shock Wave System
3 - 3
Siemens AG
RXL2-120.815.01
Page 3 of 6
LITHOSTAR Multiline
Medical Solutions
Rev. 02
02.05
CS PS 24
System manual
Filling the Coupling Circuit
3
Approximately 4 liters of distilled water are required (refer to Speed Info RX 17/96).
•
Set the following switches on board W12 (Fig. 1):
S3 in SERVICE MODE position
S2 in EMPTY COUPL. BELLOWS position (valve Y3 open)
•
Connect the test device to board W12 at test point X3.1 and connect the 0V lead to test
point X3 on board D1.
•
Check if the measured voltage of the B1 pressure meter equals1,0 V ± 0.01 V at mea-
surement point X3.1 at atmospheric pressure (adjustable with R71 on board W12).
•
Measure the zero point offset of the W4 (B3) sensor on board W12 d14
(0V on board D1X3). Record the measured value (acceptable range: 1.80 V ± 0.25 V).
•
Connect the funnel for filling the water tank (2/Fig. 2).
•
Fill the water tank until the water level is between the two marks on the upper sight glass
(1/Fig. 2).
•
Remove the funnel (open the quick coupling on the water tank) (2/Fig. 2).
•
Remove the elbow from the coupling bellows.
•
If there are air bubbles:
- Fill the hose through the elbow using a syringe and a needle.
If the hose cannot be filled, remove the shock wave head cover and fill it according to
the section "Replacing the Sensor Module" of the instructions RXL2-120.091.06..
•
Press the coupling bellows together.
•
Reconnect the elbow.
•
Set the S2 switch (Fig. 1) to STANDARD MODE (Y3 valve closed).
Fig. 1
Fig. 2
Y2
1
M20
S5
2
3
Y3
W4 (B3)
S2
S1
S3
R71
R85
R49
R998
standard mode
Empty coupl. bellows
standard mode
Service mode
Offset Coupling pressure sensor B1
Lift min. (V 22)
Lift max. (V23)
Offset Filling level sensor W4 (B3)
W12 Board
Function in
service
mode only
Button on: Filling coupling bellows