COMMISSIONING
6
85
8020776/12AX / 2019-05-31|SICK
Subject to change without notice
O P E R A T I N G I N S T R U C T I O N S |Free Flow Profiler
6.6.2.2
Calibrating the 2D LiDAR sensor
Once you have correctly aligned and adjusted the sensors, you can use the calibration
wizard to detect and correct sensor errors of measurement based on a reference object
with known dimensions.
A suitable reference object can be purchased from SICK.
Customers can also use their own reference objects with known dimensions.
Getting started
1. Place the reference object underneath the measuring gantry on the lane that you want
to monitor. Make sure that the object is located in the measuring plane of the sensors.
2. In TEMS Manager, click the
Calibration wizard
icon.
Specifying the basis of the calibration measurement
1. Go to the bottom section of the screen and click
Empty
next to the
System Component
field. From the list, choose the sensor component that you want to calibrate.
2. The
Data Point
field appears.
Note