8
FAULT DIAGNOSIS
156
8020776/12AX / 2019-05-31|SICK
O P E R A T I N G I N S T R U C T I O N S | Free Flow Profiler
Subject to change without notice
Stopping raw data recording
1. After approximately 20 minutes, switch back to editing mode and open the detail
window for the system function.
2. Set the
RawDataEnabled
parameter to
No
.
Downloading raw data via FTP
Perform an FTP download to transfer the raw data from the
/RawData
directory on the
Traffic Controller FPS.
1. Connect to the FTP server of the Traffic Controller FPS using an FTP client
(e.g., FileZilla).
2. Log in by entering
client
for the user name and
client
as the password.
Deleting raw data
Once the raw data has been downloaded from the
/RawData
directory on the Traffic
Controller FPS, you must
manually delete
the data. Otherwise, the data will clog up the
SSD by taking up unnecessary storage capacity.
The raw data are
not
deleted automatically when you deactivate the function for recording
raw data.
8.3.5
Sensor component failures
In the event of a sensor component failure, the configuration continues running. However,
the profiling system is only operable to a limited extent (see also chapter
Once a configuration affected by errors has been stopped, the profiling system can no
longer be started.
The configuration can only be restarted once the fault has been resolved.
Important
Note