13. Application Functions
13. Application Functions
13.1 Piece Counting (PCS)
1
Register PCS as one of the units beforehand.
(See
.) PCS is registered by default. There-
fore, registration is not required if the default set-
ting has not been changed.
2
From the mass display, press the
[UNIT]
key
repeatedly to switch to the PCS display.
3
When a weighing vessel (container) is used,
place the vessel on the pan and press the
[O/T]
key. (The display does not change.)
4
Count the 10 pieces (or 20, 50, 100 pieces) of
the sample to be measured accurately and load
them on the balance.
5
Press the
[CAL]
key. “Ld 10” is displayed.
6
Every time the
[CAL]
key is pressed, the dis-
play will change in this order: “Ld 10”, “Ld 20”, “Ld
50”, and “Ld 100”.
7
When the correct number of pieces on the bal-
ance is displayed and the stability mark has
appeared, press the
[O/T
] key.
8
“SEt” appears for several seconds and the
number of pieces is displayed.
9
Load the items to be measured and read the
number of pieces.
(When PCS is used for the first time)
(Example)
(When PCS is used for the first time)
(When PCS is used for the first time)
(When loading 20 pieces)
(Example)
(Example)
(Example)
(Example)
Notes
• When the
[UNIT]
key is pressed, the display returns to the mass display,
with units such as g. The actual mass of the current sample can easily
be displayed at any time this way.
• When changing the item to be counted, the process must be repeated
from step 3.
• In steps 2 through 4, the display is output in pieces based on the unit
mass of the item measured in the previous use of PCS.