11. Environment Settings
11.5 Stability Mark Lighting Timing
Stability Mark Lighting Timing can be selected according to use or demand accuracy.
1
In the gram(g) display, press
[CAL]
key several
times until
[FUnC.SEL]
is displayed.
Press
[O/T]
key.
[CAL]
is displayed.
2
Press
[CAL]
key several times until
[Stb SiGn]
is displayed.
3
Press
[O/T]
key.
[ACCUrACy]
is displayed.
Pressing
[CAL]
key alternately, the display
[ACCUrACY]
or
[FASt]
changes to another.
[ACCUrACY]
means “Accuracy”.
[FASt]
means
“Fast (Quick)”.
Stability mark
[
]
appears and means the dis-
played item has been set up now.
4
When item to be set up is displayed, press
[O/T]
key.
The item is set up and
[Stb SiGn]
is displayed.
5
Press
[POWER]
key twice to return to the
gram(g) display.
When “Stability Mark Lighting
Timing” is set to
[FASt]
A Stability Mark lights up at the same time stability is detected.
Although it becomes easy to change the measurement value after Stability
Mark lighting, since many samples are measurable one after another, the
increase in efficiency of working can be attained.
When “Stability Mark Lighting
Timing” is set to
[ACCUrACY]
After stability is detected, when the stable state continues for fixed time, a Sta-
bility Mark lights up.
Since judgment for lighting of a stability mark becomes severe and the mea-
surement value after stability mark lighting is more stabilized, it is correctly
measurable.
Fast (Quick)
Accuracy