78
13. Application Functions
(Operating Formulation Mode)
When the Formulation Mode is set on, the add-on
symbol and the Memory symbol are illuminated in the
mass display.
1
In the Formulation Mode, place the container (if
used) on the pan and press
[O/T]
key to tare. (Taring
is accepted only before weighing the first sample.)
2
Place the sample (first component) in the con-
tainer and press
[PRINT]
key.
Upon stability, the mass value is outputted to the
external Device with the numbering “CMP001”.
After output, the display is automatically tared.
3
Repeat the above step
2
until all the component
samples have been weighed.
4
Press the
[POWER]
key.
The measurements up to this point are summed
up and displayed on the balance and outputted to
the external device.
5
Clear the pan.
The balance is ready for the next set of measure-
ments.
Note
When Formulation mode is in use, fully-auto-
matic span calibration by PSC (
Clock-CAL (
) is not performed.
Instead, Weight symbol (
) keeps blink-
ing when span calibration is necessary. Span
calibration can be performed between sets
of measurements.
---Formulation Mode---
CMP001 =
0,5361g
CMP002 =
0,5422g
CMP003 =
0,4488g
TOTAL =
1,5271g
The items above are out-
putted for the measurement
examples shown here. Dec-
imal points can be either
comma or period in the out-
put. (Refer to
.)