14
4. Installation
For Basic Operation
For AUY Series
Span calibration with external weights
1
Leave the pan with nothing on, in g display
mode.
2
Press the
[CAL]
key once. “E-CAL” will be dis-
played.
3
Press the
[O/T]
key. The zero display will blink.
Stability is confirmed after about 30 seconds and
the value of the weight that should be loaded will
blink.
4
Open the glass door, load the weights of the
amount shown, and close the glass door of the
weighing chamber.
5
After a short time, the zero display will blink.
Remove the weight from the pan and close the
glass door. After “CAL End” appears, the display
returns to g display mode and the calibration is
complete.
Span calibration by external weights is set for rou-
tine calibration by default in AUY series. For other
methods, see
Note
For accurate measurement, perform calibration again when the balance is
moved or when the temperature of the installation site has changed. Daily
calibration before use is recommended.