53
10. Calibration
10.4.2 Balance ID Setting
This setting is for the balance ID number that is output along with the calibration report.
1
From mass display, press the
[CAL]
key
repeatedly until “SEttinG” appears. Press the
[O/T]
key. “CAL dEF” appears.
2
Press the
[CAL]
key repeatedly until “id:
∗∗∗∗
”
appears. (The
∗∗∗∗
is a number.) Press the
[O/T]
key. In the upper part of the display panel, the
symbol and the
symbol appear in
order to indicate numerical input mode. The left-
most digit of
∗∗∗∗
blinks. The numeral of the blink-
ing digit can be changed.
3
Press the
[UNIT]
key to increase the value of
the blinking digit by 1. Press the
[PRINT]
key to
set that digit and move the blinking digit one place
to the right. When the desired setting is entered,
press the
[O/T]
key to confirm the balance ID
number.
4
Press the
[POWER]
key to return to “id:
∗∗∗∗
”.
(No blinking)
5
Press the
[POWER]
key again to return to the
mass display.
(Example)
(Example)