69
13. Application Functions
(e) Select either the hold display mode or the con-
tinuous display mode for the specific gravity value
display. Press the
[CAL]
key repeatedly until [SG
Hold] appears.
Each time the
[O/T]
key is pressed the “
”
symbol on the left side of the display turns on or
turns off. The display is configured as shown
below according to the on/off state of “
”.
on: Hold display mode
off: Continuous display mode (the specific
gravity value is updated according to the change
of weight in water)
(f) Press the
[POWER]
key again to return to the
mass display.
5
When the
[UNIT]
key is pressed a number of
times from the mass display, [Airgd] appears for
about two seconds. Afterward, the display
switches to the [gd ] display. This is the sinker
weight measurement in air mode. During weight
measurements in air, “
” turns on in the upper-
right of the display.
6
Press the
[O/T]
key.
7
Place the sinker on the balance pan.
8
When the stability mark appears, press the
[CAL]
key. This measures the weight of the sinker
in air.
9
[wAtEr gd] appears for about two seconds.
Afterward, the display switches to the [gd ] dis-
play. This is the sinker weight measurement in liq-
uid mode. During weight measurements in liquid,
“
”turns on in the lower-right of the display.
(Example)
(Example)