66
13. Application Functions
(e) Select either the hold display mode or the con-
tinuous display mode for the specific gravity value
display. Press the
[CAL]
key repeatedly until [SG
Hold] appears.
Each time the
[O/T]
key is pressed the “
”
symbol on the left side of the display turns on or
turns off. The display is configured as shown
below according to the on/off state of “
”.
on: Hold display mode
off: Continuous display mode (the specific
gravity value is updated according to the change
of weight in water)
(f) Press the
[POWER]
key repeatedly to return to
the mass display.
5
When the
[UNIT]
key is pressed a number of
times from the mass display, [Air gd] appears for
about two seconds. Afterward, the display
switches to the [ gd ] display. This is the weight
measurement in air mode. During weight mea-
surement in air, “
” turns on in the upper-right of
the display.
6
Press the
[O/T]
key.
7
Place the items to be measured on the pan.
8
When the stability mark appears, press the
[CAL]
key. This measures the weight of the item
being measured in air.
9
The display shows [wAtEr gd] for about two
seconds. Afterward, the display switches to the
[ gd ] display. This is the weight measurement in
water mode. During weight measurement in
water, “
” turns on in the lower-right of the dis-
play.
(Example)