5. Basic Operation (Read Chapters 1 to 5 for basic but proper operation of the balance.)
For Basic Operation
Note
• Except when placing or removing items or calibration weights to and
from the weighing chamber, keep the glass doors closed unless other-
wise described in this manual.
• Air convection causes measurement error if the temperatures of the
weighed item and the chamber are different. In order to avoid this,
equalize the temperatures by leaving the item in the extra space within
the chamber before weighing.
Notes
Using a verified balance as a legal measuring instrument in the EU:
Indicates that the balance is set exactly to “Zero” with the zero-setting
function (within ±0.25e: e = verification scale interval).
Using a verified balance as a legal measuring instrument in the EU:
The balance must be used within the temperature range indicated on the
verification label.
When PSC (refer to
), fully-automatic span calibration, is not acti-
vated, operator must carry out span calibration (refer to
) upon blink-
ing of the Weight Symbol.