64
Instructions for use arco-matic 200 M / 300 M - Version 2018-05-08 GBR ID no. 2038823A
Stainless steel parts
NOTE
Only an intact stainless steel surface will not rust.
Care must therefore be taken to prevent the stain-
less steel surface from being damaged by scratches,
the use of unsuitable cleaning agents, or prolonged
contact with chlorides, acids, humidity, iron, or
similar! Always observe the following directive for
the cleaning, storage and use of stainless steel parts!
Cleaning
Only clean using standard cleaning agents for stain-
less steel products!
For mechanical cleaning, only use stainless steel
wool or brushes with natural, plastic or stainless
steel bristles! Do not use steel wool or brushes with
carbon steel!
Do not damage the surface of stainless steel prod-
ucts through scratching, scraping, etc.!
After each cleaning, always remove all cleaning
agent residues by wiping the surface thoroughly with
fresh water! Carefully dry the surface afterwards!
Retaining
Always store stainless steel products carefully
so that they are accessible to air. Always remove
cleaning fl uid residues prior to storage!
Always keep stainless steel products clean! Clean
every day to remove limescale, grease, starch and
protein layers.
Use
Do not allow stainless steel products to come into
contact with concentrated acids, acid vapours,
salt, chlorides, etc. for any longer than is absolutely
necessary!
Do not allow the stainless steel products to come
into contact with iron or steel (shavings from wires,
ferrous water) or with other metals.
Rinsing basin
The rinsing basin is made of stainless steel. Treat-
ment:
Treatment step
Treatment agent
Duration
1. Cleaning
Alkaline cleaner, temperature: 55°C
at least 10 minutes
2. Disinfection
Alkaline cleaner, temperature: 93°C
at least 10 minutes
3. Sterilisation
Steam, temperature: 134°C
at least 5 minutes
Tab. 13
Treatment steps for the rinsing basin
The values in this table are guidelines and may vary depending on the treatment device. The information
provided by the manufacturer of the treatment device applies.
Содержание 114.9600.0
Страница 83: ...Instructions for use arco matic 200 M 300 M Version 2018 05 08 GBR ID no 2038823A 83...
Страница 86: ...86 Instructions for use arco matic 200 M 300 M Version 2018 05 08 GBR ID no 2038823A...
Страница 87: ...Instructions for use arco matic 200 M 300 M Version 2018 05 08 GBR ID no 2038823A 87...
Страница 88: ...88 Instructions for use arco matic 200 M 300 M Version 2018 05 08 GBR ID no 2038823A Contact information...