18
Instructions for use arco-matic 200 M / 300 M - Version 2018-05-08 GBR ID no. 2038823A
Setting up a hardwired connection (optional)
The examination and treatment chair can, if neces-
sary, be hardwired on-site instead of connected to
the mains supply via the mains cable. This requires
some adjustments to be made in the control unit
housing.
DANGER!
Risk of electric shock!
Improper electrical work to the examination and
treatment chair puts patients, medical personnel and
those carrying out the work at risk!
The modifi cations listed in this chapter must,
therefore, only be carried out by qualifi ed specialist
personnel! The examination and treatment chair
must be completely disconnected from the mains
supply during such work!
If setting up an electrical hardwired connection for
the examination and treatment chair, a wall switch
must be installed, which can switch the examination
and treatment chair off .
Disconnect the mains plug of the examination
chair.
Undo and remove the attachment screws on the
cover of the control unit housing as shown in Fig.
11.
Remove the cover of the control unit housing as
shown in Fig. 12.
Undo the screw from the mains socket.
Undo the conductors of the mains cable on the
mains adapter.
Undo the mains cable attachment screw (Fig. 13,
Pos. 5).
Fig. 11
Remove the cover of the control unit housing
Fig. 12
Remove the cover
Fig. 13
Inside of the mains socket
1) Green and yellow protective earth conductor 2)
Brown conductor 3) Blue conductor 4) Cable attach-
ment screws 5) Mains cable attachment screw
Содержание 114.9600.0
Страница 83: ...Instructions for use arco matic 200 M 300 M Version 2018 05 08 GBR ID no 2038823A 83...
Страница 86: ...86 Instructions for use arco matic 200 M 300 M Version 2018 05 08 GBR ID no 2038823A...
Страница 87: ...Instructions for use arco matic 200 M 300 M Version 2018 05 08 GBR ID no 2038823A 87...
Страница 88: ...88 Instructions for use arco matic 200 M 300 M Version 2018 05 08 GBR ID no 2038823A Contact information...