4.2.4.4.
Changing current Active Trace to be non-ratio power Measure type
By mouse or touch screen
1) Single click [Response] and point to [Measure] in Response menu to display submenu of Measure.
2) Single click [Measure…] in submenu to display Measure dialog box.
3) Single click [More Type] to display Measure dialog box.
4) Select the source output port of Analyzer from [Source] box.
5) Click [Scale Type] to eliminate check box (if marked).
6) Select Measure receiver from Input zone.
7) Single click [OK] to close the dialog box after the setting is completed.
By front panel keys
1) Press
【
Meas
】
in RESPONSE area and single click the soft-key corresponding to [Measure…] to display Measure dialog
box.
2) Press
【
Tab
】
in ADJUST area to switch to [More Type] button and press
【
Click
】
to display Next-level Measure dialog
box.
3) Press
【
Tab
】
in ADJUST area to switch to [Source] box and press
【
↑
】
and
【
↓
】
or rotate knob to select source output
port of Analyzer.
4) Press
【
Tab
】
in ADJUST area to switch to [Scale Type] and press
【
Click
】
to eliminate check box (if marked).
5) Press
【
Tab
】
in ADJUST area to switch to Input zone and press
【
↑
】
and
【
↓
】
or rotate knob to select Measure
receiver.
6) Press
【
OK
】
in ENTRY area to close dialog box after the setting is completed.