36
Tel: 886.2.2175 2930 Email: [email protected]
By front panel keys
1) Press
【
Trace
】
key in TRACE/CHAN area.
2) Press the corresponding soft-key of [New S-Param] to create new Traces.
4.2.1.2.
Creating new Trace dialog box
Parameter select zone
It is used to select new S parameter Measure Trace to be created.
[More Type]
Opening the dialog box to create new arbitrary ratio or non-ratio power Measure Traces.
[Ch Select]
It is used for selecting the channel for a new Trace.
[Create in New Window]
A Trace is created in a new window when it is checked and is generated in the current window at the elimination.
4.2.2.
Arbitrary ratio Measure
The arbitrary ratio Measure allows to select input signals and reference signals for ratio Measure from receivers (A, B, R1 and R2).
4.2.2.1.
Creating new arbitrary ratio Measure Trace
By mouse or touch screen
1) Single click [Trace/Chan], single click [New Trace...] in [Trace] menu, and display the dialog box of Create New Trace.
2) Single click [More Type] button in the dialog box to display the dialog box of New Trace.
3) Select the output port of Analyzer source signals from [Source] box.
4) Click [Scale Type] to mark check box.
5) Select input receiver for ratio Measure from Input zone.
6) Select reference receiver for ratio Measure from Reference zone.
7) Single click [OK] button to close the dialog box.
Fig. 4-4 Creating New Arbitrary Ratio Measure Trace