110
Tel: 886.2.2175 2930 Email: [email protected]
6.8.
Parameter Transformation
Parameter transformation function can transform the test result of S parameter into other parameters in real time.
6.8.1.
Parameter transformation
6.8.1.1.
By mouse or touch screen
1) Click [Marker/Analysis] menu
,
and click [Conversion] on the menu of [Analysis] to open the dialog box of parameter
transformation.
2) Click the [Enable] check box in the dialog box to enable the parameter transform function.
3) Click the parameter transformation type box in the dialog box and select the parameter type to be transformed on
the drop-down menu.
Figure 6-26 Parameter Transformation Dialog Box
6.8.1.2.
By front panel keys
1) Press
【
Analysis
】
in MKR/ANALYSIS area and press [Conversion] in the soft-key menu that turns up.
2) Press [Conversion on|OFF] to trigger the opening and closing of parameter transformation function.
3) Press [Function], and select corresponding transformation type in the sub-menu of soft-key that turns up.
6.8.2.
Function type of parameter transformation
Parameter transformation function is used to transform the test result of S parameter
(
Sxy) into the following parameters:
Impedance (
r
Z
) and admittance (
r
Y
) in reflection test
xy
xy
0X
r
S
1
S
1
Z
Z
,
r
r
Z
1
Y
Impedance(
t
Z
)and admittance(
t
Y
) in transmission test
0Y
0X
xy
0Y
0X
t
Z
Z
S
Z
Z
2
Z
,
t
t
Z
1
Y
Impedance(
s
Z
t
)and admittance (
ts
Y
) in the parallel transmission