Fig.8-21 Setting Time Domain Transformation Measure
e) Set Window function: click [Window] Selection Button in Time Domain Transform dialog box and set window filter with
three methods below:
Drag sliding box with mouse till pulse width or sidelobe level meets the requirement.
Click
[Gaesarian Window β Para]
to adjust setting of value β, till pulse width or sidelobe level meets the requirement.
Click
[Pulse Span]
to set response pulse width directly.
Fig.8-22 Time Domain Transformation Measure—Setting Window
f)
Set filtering function of time domain gate: click [Gate Setting] option button in Time Domain Transform dialog box and
click to select [Gating] check box to enable filtering function of time domain gate.
Click [Start], [Stop] or [Center], [Span] box to set gate border.
Select gate type in [Gate Type] box: band pass, band stop.
Select gate form in [Gate Shape]: minimum, normal, wide, maximum.