96
Tel: 886.2.2175 2930 Email: [email protected]
5) Press corresponding soft-key of [Fail Sign on/OFF] to trigger display of FAIL Mark of limit test.
6) Press corresponding soft-key of [Edit Limit Table] , and press corresponding soft-key of [Limit Table on/OFF] on the
sub-menu to trigger display of limit table.
Limit test is only made at actual Sweep measured point, when Sweep points are less, performance index of measured piece may not
meet requirements, but it still can pass limit test, therefore during actual Measure, adequate canning points must be used to make
limit test.
6.3.2.3.
Limit Test dialog box
[Limit Test(ON/off)]
When it is selected, limit test function of Active Trace will be enabled, when it is cleared, limit test function of Active Trace will be
closed.
[Limit Line(ON/off)]
When it is selected, limit line of Active Trace will be displayed on screen, when it is cleared, display of limit line of Active Trace will be
closed, but it will not affect limit test function.
[Sound on Fail]
When it is selected, if Trace data point test is failed, buzzer will send out warning tone.
[FAIL Mark(ON/off)]
When it is selected, if Trace data point test is failed, the mark of fail will be displayed on the screen.
[Dot]
Use symbol “v” (Measure type as MAX) and “^”(Measure type as MIN)to indicate corresponding discrete limit value of measured data
point during selecting.
[Line]
Use line to connect all discrete limit set point during selecting.
[Show Table]
When it is clicked, limit table will be showed for edition, and when the limit table is enabled, the button will change into forbidden
status.
[Hide Table]
When it is clicked, limit table display will be closed, and if it is closed, the button will change into forbidden status.
[Export Table]
When it is clicked, storage dialog box will turn up, you can select storage route and store the file with *.csv format.
[Import Table]
When it is clicked, storage dialog box will turn up, select the storage route of the file to be opened and you can open file with *.csv
format.
[Limit Test Type]
Single-click the drop-down box, you can select limit test type. There are totally four types: Multi-type limit line
、
Declined limit line
、