112
Tel: 886.2.2175 2930 Email: [email protected]
Fig.6-27 Time Domain Transformation Measure—Setting Window
6) Set filtering function of time domain gate: click [Gate Setting] option button in Time Domain Transform dialog box and click
to select [Gating] check box to enable filtering function of time domain gate.
Click [Start], [Stop] or [Center], [Span] box to set gate border.
Select gate type in [Gate Type] box: band pass, band stop.
Select gate form in [Gate Shape]: minimum, normal, wide, maximum.
Fig.6-28 Time Domain Transformation Measure—Gate Setting
7) If it is necessary to observe frequency domain response of the measured piece when the gate function is still effective,
click [Transform] option button in Time Domain Transform dialog box and click to clear [Transform] check box.
8) Click [OK] to close Time Domain Transform dialog box after setting.
6.9.2.
By front panel keys—Time Domain Transform Dialog Box
1) Press
【
Analysis
】
in MKR/ANALYSIS area, press soft-key corresponding to [Transform], relative soft-key menu appears,
press corresponding soft-key of [Transform …] to display Time Domain Transform dialog box.
2) Press
【
Tab
】
in ADJUST area to switch to [Transform] check box, press
【
Click
】
to select check box to Stimulus time
domain transformation function.
3) Press
【
Tab
】
to switch to Transformation Mode zone, Press
【
↑
】
,
【
↓
】
in ADJUST area or rotate knob to select [Low
Pass Impulse], [Low Pass Step] or [Band Pass] mode.
4) Press
【
Tab
】
to switch to [Start], [Stop] or [Center], [Span] box to set time domain Measure range.
5) Set window function: press
【
Tab
】
to switch to [Transform] option button, press
【
↑
】
,
【
↓
】
or rotate knob to switch
to [Window] option button and then set window filter in three methods below: