90
Reference Manual
00809-0200-4840, Rev BA
Section 5: Service and Troubleshooting
December 2014
Section 5: Service and Troubleshooting
5.5.7
Distance (D910)
This indicates the distance from the transmitter face to a detected surface.
It is a useful diagnostic because a false surface signal can be identified and related to the physical
nature of the installation.
Note
shows this parameter in a tank geometry illustration, and how
Distance Offset (
) does not affect this value.
Field Communicator or AMS
To view the distance to a detected surface:
1.
From the
Home
screen, select 3: Service Tools.
2.
Select 1: Alerts.
3.
Select 2: Diagnostic Data.
4.
Select 1: Target Data.
5.
Select 1: Distance D910.
6.
When finished, press EXIT to exit to the previous menu.
Rosemount 3490 Series Control Unit
To view the distance to a detected surface:
1.
From the
Main Menu
screen, select MONITOR.
2.
Select the transmitter (e.g. “
Tx1: 3107
”).
3.
Select DIAGNOSTICS.
4.
Select Distance.
5.
When finished, press the ESC key to exit to the previous menu.
Fast Keys
3, 1, 2, 1, 1
Distance D910
2.327 ft
EXIT
HELP
(Field Communicator Screen)
Distance D910
2.227ft
(Rosemount 3491 Screen)
Содержание 3107
Страница 2: ......
Страница 4: ...Reference Manual 00809 0200 4840 Rev BA Title Page December 2014 ...
Страница 10: ...TOC vi Reference Manual 00809 0200 4840 Rev BA Table of Contents December 2014 Table of Contents ...
Страница 16: ...6 Reference Manual 00809 0200 4840 Rev BA Section 2 Transmitter Overview December 2014 ...
Страница 90: ...80 Reference Manual 00809 0200 4840 Rev BA Section 4 Configuration December 2014 Section 4 Configuration ...
Страница 144: ...134 Reference Manual 00809 0200 4840 Rev BA Appendix D Field Communicator December 2014 Appendix D Field Communicator ...
Страница 145: ......
Страница 146: ......
Страница 147: ......